Current Directions in Biomedical Engineering (Sep 2019)

Cytotoxicity of Catalysed Silicone Resin Coatings for Smart Biomedical Devices

  • Werner Valerie M. K.,
  • Kist Arian,
  • Eblenkamp Markus

DOI
https://doi.org/10.1515/cdbme-2019-0042
Journal volume & issue
Vol. 5, no. 1
pp. 165 – 170

Abstract

Read online

Equipping medical devices with smart technologies holds great potential for the development of modern medical products. The development requires the identification of new integration strategies and the research of new material combinations due to the miniaturization of systems and increasing production figures. The realization of Smart Biomedical Devices requires a sufficient barrier effect (bioprotection) by appropriate encapsulation of the electronic components. Thinnest polymer coatings have proven to be suitable for conformal encapsulation. The aim of the study was to investigate the fundamental suitability of thin-film lacquers added with catalysts as coating materials for electronic systems with regard to their biological use. Due to long curing times of up to 14 days, eight different catalysts based on different chemical structures were added to the coating materials and their influence on a cytotoxic effect was investigated. A non-cytotoxic effect was observed for the organometallic catalysts based on tin, zirconium, titanium, bismuth, and tertiary amine. Most were resistant to steam sterilization. The curing time of the non-cytotoxic coatings could be significantly reduced by the addition of catalysts. The shortening of process times is an important economic aspect in the production of mass-produced Smart Biomedical Devices.

Keywords