Arthroplasty Today (Oct 2022)
Tibial Baseplate-Cement Interface Debonding in the ATTUNE Total Knee Arthroplasty System
Abstract
Background: Acrylic bone cement is the most common method of fixation for primary total knee arthroplasty (TKA). Several studies have described good short-term outcomes; however, there have been reports of early failures due to tibial baseplate debonding at the implant-cement interface of The ATTUNE Knee System (DePuy Synthes, West Chester, PA). We examined the causes and rates of revision in patients who underwent TKA with this system to identify factors associated with this mode of early failure. Methods: A retrospective review of electronic health records between 2013 and 2018 identified all patients undergoing TKA with the ATTUNE Knee System with a minimum 2-year follow-up. Cause of revision, patient, implant, instrumentation, cement, and surgeon variables were collected. A descriptive analysis was used to identify characteristics of surgeon (fellowship-trained, surgical volume), implant (baseplate, bearing), and cement (brand, viscosity) that were associated with aseptic loosening. Results: A total of 668 patients representing 742 knees were identified. Eighteen (2.4%) required a revision surgery. Aseptic loosening was the leading cause of revision surgery (n = 10, 55.6%). All failures due to aseptic loosening involved debonding of the tibial implant-cement interface. A multivariate analysis identified low-volume surgeons (9.0%, P < .0001) and 1 specific brand of high-viscosity cement (14.3%, P < .0001) as risk factors for aseptic loosening. Conclusions: This study represents the largest nonregistry review of the original ATTUNE Knee System. Surgeon case volume and cement viscosity were factors associated with an increased rate of early failure due to tibial baseplate implant-cement interface debonding.