Cardiovascular Ultrasound (Oct 2020)

Three-dimensional transthoracic echocardiographic evaluation of tricuspid regurgitation severity using proximal isovelocity surface area: comparison with volumetric method

  • Beiqi Chen,
  • Yu Liu,
  • Wuxu Zuo,
  • Quan Li,
  • Dehong Kong,
  • Cuizhen Pan,
  • Lili Dong,
  • Xianhong Shu,
  • Junbo Ge

DOI
https://doi.org/10.1186/s12947-020-00225-y
Journal volume & issue
Vol. 18, no. 1
pp. 1 – 12

Abstract

Read online

Abstract Background The quantification of tricuspid regurgitation(TR) using three-dimensional(3D) proximal isovelocity surface area (PISA) derived effective regurgitant orifice area (EROA) is feasible in functional TR. The aim of our study was to explore the diagnostic accuracy and utility of 3D PISA EROA in a larger population of different etiologies. Methods One hundred and seven patients with confirmed TR underwent 2D and 3D transthoracic echocardiography (TTE). 3D PISA EROA was calculated and EROA derived from 3D regurgitant volume (Rvol) was used as the reference. Results 3D PISA EROA showed better correlation in primary TR than in functional TR(r = 0.897, P < 0.01). 3D PISA EROA differentiated severe TR with comparable accuracy in patients with primary and functional etiology (Z-value 16.506 vs 21.202), but with different cut-offs (0.49cm2 vs. 0.41 cm2). The chi-square value for incorporated clinical symptoms, positive echocardiographic results and 3D PISA EROA to grade severe TR was higher than only included clinical symptoms or incorporated clinical symptoms and positive echocardiographic results (chi-square value 137.233, P < 0.01). Conclusion TR quantification using 3D PISA EROA is feasible and accurate under different etiologies. It has incremental diagnostic value for evaluating severe TR.

Keywords