A Novel Non-Invasive Thermometer for Continuous Core Body Temperature: Comparison with Tympanic Temperature in an Acute Stroke Clinical Setting
Miloš Ajčević,
Alex Buoite Stella,
Giovanni Furlanis,
Paola Caruso,
Marcello Naccarato,
Agostino Accardo,
Paolo Manganotti
Affiliations
Miloš Ajčević
Clinical Unit of Neurology, Department of Medicine, Surgery and Health Sciences, Cattinara University Hospital ASUGI, University of Trieste, Strada di Fiume, 447-34149 Trieste, Italy
Alex Buoite Stella
Clinical Unit of Neurology, Department of Medicine, Surgery and Health Sciences, Cattinara University Hospital ASUGI, University of Trieste, Strada di Fiume, 447-34149 Trieste, Italy
Giovanni Furlanis
Clinical Unit of Neurology, Department of Medicine, Surgery and Health Sciences, Cattinara University Hospital ASUGI, University of Trieste, Strada di Fiume, 447-34149 Trieste, Italy
Paola Caruso
Clinical Unit of Neurology, Department of Medicine, Surgery and Health Sciences, Cattinara University Hospital ASUGI, University of Trieste, Strada di Fiume, 447-34149 Trieste, Italy
Marcello Naccarato
Clinical Unit of Neurology, Department of Medicine, Surgery and Health Sciences, Cattinara University Hospital ASUGI, University of Trieste, Strada di Fiume, 447-34149 Trieste, Italy
Agostino Accardo
Department of Engineering and Architecture, University of Trieste, Via A. Valerio, 10-34127 Trieste, Italy
Paolo Manganotti
Clinical Unit of Neurology, Department of Medicine, Surgery and Health Sciences, Cattinara University Hospital ASUGI, University of Trieste, Strada di Fiume, 447-34149 Trieste, Italy
There is a growing research interest in wireless non-invasive solutions for core temperature estimation and their application in clinical settings. This study aimed to investigate the use of a novel wireless non-invasive heat flux-based thermometer in acute stroke patients admitted to a stroke unit and compare the measurements with the currently used infrared (IR) tympanic temperature readings. The study encompassed 30 acute ischemic stroke patients who underwent continuous measurement (Tcore) with the novel wearable non-invasive CORE device. Paired measurements of Tcore and tympanic temperature (Ttym) by using a standard IR-device were performed 3–5 times/day, yielding a total of 305 measurements. The predicted core temperatures (Tcore) were significantly correlated with Ttym (r = 0.89, p p = 0.923). The Tcore measurements correctly predicted the presence or absence of Ttym hyperthermia or fever in 94.1% and 97.4% of cases, respectively. Temperature monitoring with a novel wireless non-invasive heat flux-based thermometer could be a reliable alternative to the Ttym method for assessing core temperature in acute ischemic stroke patients.