BMC Musculoskeletal Disorders (Jul 2024)
Screw placement through a higher medial portal provides better initial stability in arthroscopic ACL tibial avulsion fracture fixation: a finite element analysis
Abstract
Abstract Objective The objective of this study was to investigate the initial stability of different screw placements in arthroscopic anterior cruciate ligament (ACL) tibial avulsion fracture fixation. Methods A three-dimensional knee model at 90° flexion was utilized to simulate type III ACL tibial avulsion fracture and arthroscopic screw fixation through different portals, namely the central transpatellar tendon portal (CTP), anterolateral portal (ALP), anteromedial portal (AMP), lateral parapatellar portal (LPP), medial parapatellar portal (MPP), lateral suprapatellar portal (LSP), medial suprapatellar portal (MSP). A shear force of 450 N was applied to the finite element models at 30° flexion to simulate the failure condition. The displacement of the bony fragment and the volume of the bone above 25,000 µ-strain (damaged bone volume) were calculated around the screw path. Results When the screw was implanted through CTP, the displacement of the bony fragment reached the maximum displacement which was 1.10 mm and the maximum damaged bone volume around the screw path was 148.70 mm3. On the other hand, the minimum displacement of the bony fragment was 0.45 mm when the screw was implanted through LSP and MSP. The minimum damaged bone volume was 14.54 mm3 around the screw path when the screw was implanted through MSP. Conclusion Screws implanted through a higher medial portal generated less displacement of the bony fragment and a minimum detrimental strain around the screw path. The findings are clinically relevant as they provide biomechanical evidence on optimizing screw placement in arthroscopic ACL tibial avulsion fracture fixation.
Keywords