Practical Laboratory Medicine (Jan 2024)
Performance evaluation of the LumiraDx quantitative microfluidic point-of-care CRP test
Abstract
C-reactive protein (CRP) is an established acute-phase marker for infection, inflammation and tissue injury, used to guide clinical decision-making in primary and secondary care. This study compared the analytical performance of the quantitative microfluidic point-of-care LumiraDx CRP Test to a laboratory-based reference method (Siemens RCRP Flex assay on the Dimension® Xpand®) and evaluated equivalence of sample matrices (blood versus plasma) in point-of-care settings using samples from patients presenting with symptoms of infection or inflammation. The LumiraDx CRP Test demonstrated close agreement with the lab reference test (range, 5.1 to 245.2 mg/L, r = 0.992, slope = 0.998, intercept = –0.476; n = 205) and notable agreement between fingerstick and venous blood and plasma (r = 0.974–0.983; n = 44). Paired replicate precision had mean coefficients of variation of 6.4 % (plasma), 6.6 % (capillary direct) and 8.1 % (venous blood); overall error rates were 2.9 %. The quantitative LumiraDx CRP Test showed robust analytical performance across sample matrices and close agreement compared to the laboratory reference method when used at the point of care.