Materials (Aug 2020)

Retentive Characteristics of a New Attachment System for Hybrid Dentures

  • Christin Arnold,
  • Charlotte Stampa,
  • Ramona Schweyen,
  • Jeremias Hey,
  • Arne Boeckler

DOI
https://doi.org/10.3390/ma13153434
Journal volume & issue
Vol. 13, no. 15
p. 3434

Abstract

Read online

Removable implant-anchored dentures have become an established treatment concept especially for older, multimorbid patients. This study investigates the retention force (RF) of two different attachment systems. A total of 96 specimens (n = 8 for each condition) were fabricated and RF was measured under different conditions: fatigue (10,000 cycles dislodging), thermal undulation (5/55 °C, 5000 cycles) and implant-angulation (0°, 5°, 10°, 15°, 20°). The Novaloc system ((N), 0° and 15° abutments, yellow matrix (Y)) was compared to the Locator system ((L), pink (P) and orange (O)). Initial RFs (8.57 ± 0.99 N (NY), 19.39 ± 8.10 N (LP), 8.8 ± 5.28 N (LO)) were reduced by ageing simulation (26% (NY), 66% (LP), 89% (LO); p p p = 0.002, LO: p = 0.148). In contrast to LP, the RFs of Novaloc abutments and LOs predominantly showed no clinically relevant dependence on implant angulation. Ageing processes tended to result in lower RFs at higher implant angulation. Thus, the Novaloc attachment system offers an alternative to Locator attachments. It is characterized by a comparatively continuous RF-curve over the entire wearing period. Future clinical studies have to be conducted to verify the in vitro demonstrated advantages of the Novaloc system.