Polymers (Sep 2021)

Development, Characterization and Valuable Use of Novel Dosimeter Film Based on PVA Polymer Doped Nitro Blue Tetrazolium Dye and AgNO<sub>3</sub> for the Accurate Detection of Low X-ray Doses

  • Saleh Alashrah,
  • Yassine El-Ghoul,
  • Faisal Muteb Almutairi,
  • Mohammed Ahmed Ali Omer

DOI
https://doi.org/10.3390/polym13183140
Journal volume & issue
Vol. 13, no. 18
p. 3140

Abstract

Read online

Currently, the uncontrolled exposure of individuals to X-rays during medical examinations represents a substantial danger that threatens both medical professionals and patients. Therefore, radiation dosimetry for low X-ray doses is a very important control of radiation practice in medical diagnostic radiology. In line with this, the current study proposes a valuable dosimeter-based PVA thin film doubly doped with silver nitrate salt and nitro blue tetrazolium dye. The nanocomposite film was prepared via a simple casting method and the different processing parameters were optimized. The performance of radiation detection was evaluated according to optical, chromic, chemical and structural changes after exposure to variable low X-ray doses (0, 2, 4, 10 and 20 mGy). The different film labels exhibited an excellent stability behavior in dark and light upon 30 days of storage. The UV-Vis spectrophotometric study showed a gradual increase in the maximum absorbance as a function of the dose and the corresponding response curve confirmed this linear variation (R = 0.998). A clear structural modification was recorded via X-ray diffraction (XRD) analysis revealing the increase in crystallinity with the level of the dose received by the nanocomposite films. Microscopic surface analysis via SEM assessments revealed a significant morphological change in PVA/Ag+/NBT films exposed to increased radiation doses and typical dendrites growing in needle- or tree-like microstructures appeared with a high X-ray dose. Finally, the nanocomposite films before and after irradiation were evaluated via a spectrocolorimetric study and the different CIELab coordinates, the color difference, as well as the color strength, showed a linear correlation with the intensity of the applied dose. This new dosimeter design could, therefore, provide a promising and efficient alternative for prompt and accurate detection of low X-rays doses in diagnostic radiology.

Keywords