Sensors (Oct 2021)

Heart and Lung Sound Measurement Using an Esophageal Stethoscope with Adaptive Noise Cancellation

  • Nourelhuda Mohamed,
  • Hyun-Seok Kim,
  • Kyu-Min Kang,
  • Manal Mohamed,
  • Sung-Hoon Kim,
  • Jae Gwan Kim

DOI
https://doi.org/10.3390/s21206757
Journal volume & issue
Vol. 21, no. 20
p. 6757

Abstract

Read online

In surgeries where general anesthesia is required, the auscultation of heart and lung sounds is essential to provide information on the patient’s cardiorespiratory system. Heart and lung sounds can be recorded using an esophageal stethoscope; however, there is huge background noise when this device is used in an operating room. In this study, a digital esophageal stethoscope system was designed. A 3D-printed case filled with Polydimethylsiloxane material was designed to hold two electret-type microphones. One of the microphones was placed inside the printed case to collect the heart and lung sound signals coming out from the patient through the esophageal catheter, the other was mounted on the surface of the case to collect the operating room sounds. A developed adaptive noise canceling algorithm was implemented to remove the operating room noise corrupted with the main heart and lung sound signals and the output signal was displayed on software application developed especially for this study. Using the designed case, the noise level of the signal was reduced to some extent, and by adding the adaptive filter, further noise reduction was achieved. The designed system is lightweight and can provide noise-free heart and lung sound signals.