Sensors (Dec 2018)

Scale Factor Calibration for a Rotating Accelerometer Gravity Gradiometer

  • Zhongguang Deng,
  • Chenyuan Hu,
  • Xiangqing Huang,
  • Wenjie Wu,
  • Fangjing Hu,
  • Huafeng Liu,
  • Liangcheng Tu

DOI
https://doi.org/10.3390/s18124386
Journal volume & issue
Vol. 18, no. 12
p. 4386

Abstract

Read online

Rotating Accelerometer Gravity Gradiometers (RAGGs) play a significant role in applications such as resource exploration and gravity aided navigation. Scale factor calibration is an essential procedure for RAGG instruments before being used. In this paper, we propose a calibration system for a gravity gradiometer to obtain the scale factor effectively, even when there are mass disturbance surroundings. In this system, four metal spring-based accelerometers with a good consistency are orthogonally assembled onto a rotary table to measure the spatial variation of the gravity gradient. By changing the approaching pattern of the reference gravity gradient excitation object, the calibration results are generated. Experimental results show that the proposed method can efficiently and repetitively detect a gravity gradient excitation mass weighing 260 kg within a range of 1.6 m and the scale factor of RAGG can be obtained as (5.4 ± 0.2) E/μV, which is consistent with the theoretical simulation. Error analyses reveal that the performance of the proposed calibration scheme is mainly limited by positioning error of the excitation and can be improved by applying higher accuracy position rails. Furthermore, the RAGG is expected to perform more efficiently and reliably in field tests in the future.

Keywords