Bangladesh Journal of Pharmacology (Jul 2017)

Estimation of vitamin B12 in plasma by High Performance Liquid Chromatography

  • Fahmida Hasnat,
  • Habiba Akhter Bhuiyan,
  • Mir Misbahuddin

DOI
https://doi.org/10.3329/bjp.v12i3.33108
Journal volume & issue
Vol. 12, no. 3

Abstract

Read online

High performance liquid chromatography was used to develop and validate the detection of vitamin B12 in blood plasma sample. The mobile phase consists of a mixture of 0.025% trifluroacetic acid in deionized water and 30% methanol. The mixture used was adjusted to pH 2.9 and the flow rate was adjusted to 0.5 mL/min. The separation was achieved using C18 column maintained at 30ºC temperature and detection of vitamin B12 was conducted at maximum wavelength 230 nm. Video Clip of Methodology: Estimation of vitamin B12 in plasma by HPLC: 4 min 29 sec Full Screen Alternate

Keywords