Journal of Intensive Care (Aug 2025)

Preliminary application of the Sharesource connect data collection and analysis system in the management of continuous renal replacement therapy in the intensive care unit

  • Wensan Wu,
  • Jianli Wang,
  • Chen Chen,
  • Junqi Feng,
  • Shuyi Zhang,
  • An Shi,
  • Jing Zhang,
  • Xinyi He,
  • Jiangchen Peng,
  • Mingli Zhu

DOI
https://doi.org/10.1186/s40560-025-00818-7
Journal volume & issue
Vol. 13, no. 1
pp. 1 – 11

Abstract

Read online

Abstract Background This study investigated the application of software-based data monitoring for quality control (QC) in continuous renal replacement therapy (CRRT) management. Methods This single-center pre-post intervention study, conducted in three ICUs of a tertiary hospital in Shanghai, compared outcomes before (Jan–Dec 2023) and after (Jan–Dec 2024) implementing the Sharesource Connect system. Data from 9 Prismaflex CRRT machines were collected retrospectively during 2023 and prospectively on a monthly basis during 2024. Alongside the software, a comprehensive quality improvement program: (1) multidisciplinary team collaboration; (2) data-driven QC; and (3) structured training. Primary outcomes—filter lifespan, downtime percentage, delivered/prescribed dose ratio, ultrafiltration volume, and vascular access alarms—were compared. Results A total of 798 filters from 514 patients (2023) and 717 filters from 492 patients (2024) were analyzed. Key quality metrics improved significantly following implementation (2024 vs. 2023): Filter lifespan increased significantly from 20.08 ± 4.12 h to 24.08 ± 4.27 h (P = 0.043), Kaplan–Meier analysis demonstrated improved filter survival (Log-Rank p < 0.001). Cumulative survival increased from 2023 to 2024 at key time points: 12 h (69.1%–87.2%, + 18.1%), 24 h (30.9%–34.6%, + 3.7%), and 36 h (5.6%–13.6%, + 8.0%), with consistent improvements observed. Downtime percentage decreased from 39 to 28% (P = 0.015), reducing non-effective treatment time by 11 percentage points. The delivered/prescribed dose ratio increased from 82 to 86% (P = 0.046). The mean delivered dose was 35.67 ± 4.01 mL/kg/h (prescribed: 41.33 ± 4.5 mL/kg/h). Ultrafiltration volume remained stable (3.13 ± 0.37 vs. 3.52 ± 0.44 L/treatment day, P = 0.058). There was no significant difference in vascular access alarms (3.39 ± 1.44 vs. 2.93 ± 0.73 events/day, P = 0.392). Conclusion The Sharesource Connect system could be used for the monitoring, collection, and analysis of CRRT data to assist in the QC management related to CRRT, so as to provide a software basis for further multi-center studies or random control trials on the intelligent management of critical patients undergoing CRRT.

Keywords