Tremor and Other Hyperkinetic Movements (Oct 2019)

Position-Dependent Dysfunction of Deep Brain Stimulation in Tourette Syndrome: Diagnostic Clues

  • Marc E. Wolf,
  • Olaf Majewski,
  • Kirsten R. Müller-Vahl,
  • Christian Blahak,
  • Dirk-Michael Schulte,
  • Joachim K. Krauss

DOI
https://doi.org/10.7916/tohm.v0.713
Journal volume & issue
Vol. 9, no. 0
pp. 1 – 3

Abstract

Read online

Background: Detection of defective deep brain stimulation (DBS) contacts/electrodes is sometimes challenging. Case Report: We report a patient with Tourette syndrome (TS), who presented with abrupt tic increase and mild generalized headache 9 years after DBS implantation. On the suspicion of a hardware defect, a fracture of the DBS electrode and extension lead was ruled out by radiography and standard implantable pulse generator readouts. Further investigation revealed position-dependent modifiable therapeutic impedances, suggesting an impaired contact of the extension lead/adaptor. After replacement normal impedances were recorded, and the patient fully recovered. Discussion: In DBS dysfunction with inconspicuous hardware check, position-dependent defects might be suspected.

Keywords