European Radiology Experimental (Apr 2021)

Influence of beam hardening in dual-energy CT imaging: phantom study for iodine mapping, virtual monoenergetic imaging, and virtual non-contrast imaging

  • Risa Kanatani,
  • Takashi Shirasaka,
  • Tsukasa Kojima,
  • Toyoyuki Kato,
  • Masateru Kawakubo

DOI
https://doi.org/10.1186/s41747-021-00217-1
Journal volume & issue
Vol. 5, no. 1
pp. 1 – 8

Abstract

Read online

Abstract In this study, we investigated the influence of beam hardening on the dual-energy computed tomography (DECT) values of iodine maps, virtual monoenergetic (VME) images, and virtual non-contrast (VNC) images. 320-row DECT imaging was performed by changing the x-ray tube energy for the first and second rotations. DECT values of 5 mg/mL iodine of the multi-energy CT phantom were compared with and without a 2-mm-thick attenuation rubber layer (~700 HU) wound around the phantom. It was found that the CT density values UH, with/without the rubber layer had statistical differences in the iodine map (184 ± 0.7 versus 186 ± 1.8), VME images (125 ± 0.3 versus 110 ± 0.4), and VNC images (−58 ± 0.7 versus −76 ± 1.7) (p < 0.010 for all). This suggests that iodine mapping may be underestimated by DECT and overestimated by VME imaging because of x-ray beam hardening. The use of VNC images instead of plain CT images requires further investigation because of underestimation.