The Journal of Indian Prosthodontic Society (Jan 2022)
Retention force of Molloplast-B with ball attachment in implant-supported overdentures: An in vitro study
Abstract
Aim: The purpose of this study is to evaluate the retention effectiveness of Molloplast B as a female attachment compared to O rings' in implant supported overdentures. Settings and Design: This systematic review and meta-analysis was evaluated using the Preferred Reporting Items for Systematic Review and Meta-Analysis guidelines. Materials and Methods: Sixteen female part models were divided into two groups: eight female parts made with O ring (Group A) and eight female parts made with Molloplast B (Group B). All of the models were soaked in artificial saliva for 24 h, then, their retention force was measured in Newton using a Universal mechanical testing machine, initially, after 500, after 1000, and after 1500 of loading and dislodging cycles. Statistical Analysis Used: The statistical analysis was conducted by using one way ANOVA test and Bonferroni test. SPSS Software (SPSS, Version 27, IBM Co., Chicago, IL, USA). Results: After 1500 loading and dislodging cycles, Group B has the highest mean retention force (4.09), followed by Group A, which has a mean retention force of 3.73. Conclusion: Molloplast B with a 2.7 mm diameter ball attachment lost the least amount of retention force after 1500 loading and dislodging cycles.