Clinical Case Reports (Dec 2022)

A ruptured balloon shaft during an angioplasty

  • Rania Hammami,
  • Tarek Ellouze,
  • Amine Bahloul,
  • Leila Abid

DOI
https://doi.org/10.1002/ccr3.6600
Journal volume & issue
Vol. 10, no. 12
pp. n/a – n/a

Abstract

Read online

Abstract We report a case of balloon shaft rupture during percutaneous coronary intervention. Although the entrapped balloon was not yet deflated when the complication occurred, we successfully retrieved it percutaneously using a trapping technique. This case described a cheap and straightforward technique of device retrieval that helped save our patient.

Keywords