PLoS ONE (Jan 2023)

A computer vision based optical method for measuring fluid level in cell culture plates.

  • Pierre V Baudin,
  • Mircea Teodorescu

DOI
https://doi.org/10.1371/journal.pone.0290951
Journal volume & issue
Vol. 18, no. 9
p. e0290951

Abstract

Read online

For a transparent well with a known volume capacity, changes in fluid level result in predictable changes in magnification of an overhead light source. For a given well size and fluid, the relationship between volume and magnification can be calculated if the fluid's index of refraction is known or in a naive fashion with a calibration procedure. Light source magnification can be measured through a camera and processed using computer vision contour analysis with OpenCV. This principle was applied in the design of a 3D printable sensing device using a raspberry pi zero and a camera.