European Journal of Radiology Open (Jan 2018)

Image quality with iterative reconstruction techniques in CT of the lungs—A phantom study

  • Hilde Kjernlie Andersen,
  • David Völgyes,
  • Anne Catrine Trægde Martinsen

Journal volume & issue
Vol. 5
pp. 35 – 40

Abstract

Read online

Background: Iterative reconstruction techniques for reducing radiation dose and improving image quality in CT have proved to work differently for different patient sizes, dose levels, and anatomical areas. Purpose: This study aims to compare image quality in CT of the lungs between four high-end CT scanners using the recommended reconstruction techniques at different dose levels and patient sizes. Material and methods: A lung phantom and an image quality phantom were scanned with four high-end scanners at fixed dose levels. Images were reconstructed with and without iterative reconstruction. Contrast-to-noise ratio, modulation transfer function, and peak frequency of the noise power spectrum were measured. Results: IMR1 Sharp+ and VEO improved contrast-to-noise ratio to a larger extent than the other iterative techniques, while maintaining spatial resolution. IMR1 Sharp+ also maintained noise texture. Conclusions: IMR1 Sharp+ was the only reconstruction technique in this study which increased CNR to a large extent, while maintaining all other image quality parameters measured in this study. Keywords: CT, Lung, Physics, Iterative reconstruction