Journal of Medical Physics (Jan 2007)
Initial experience with an 11 MeV self-shielded medical cyclotron on operation and radiation safety
Abstract
A self-shielded medical cyclotron (11 MeV) was commissioned at our center, to produce positron emitters, namely, 18 F, 15 O, 13 N and 11 C for positron emission tomography (PET) imaging. Presently the cyclotron has been exclusively used for the production of 18 F - for 18 F-FDG imaging. The operational parameters which influence the yield of 18 F - production were monitored. The radiation levels in the cyclotron and radiochemistry laboratory were also monitored to assess the radiation safety status in the facility. The target material, 18 O water, is bombarded with proton beam from the cyclotron to produce 18 F - ion that is used for the synthesis of 18 F-FDG. The operational parameters which influence the yield of 18 F - were observed during 292 production runs out of a total of more than 400 runs. The radiation dose levels were also measured in the facility at various locations during cyclotron production runs and in the radiochemistry laboratory during 18 F-FDG syntheses. It was observed that rinsing the target after delivery increased the number of production runs in a given target, as well as resulted in a better correlation between the duration of bombardment and the end of bombardment 18 F - activity with absolutely clean target after being rebuilt. The radiation levels in the cyclotron and radiochemistry laboratory were observed to be well within prescribed limits with safe work practice.