Frontiers in Biomedical Technologies (Dec 2021)

Measurement and Calculation of SAR Due to 900 MHz Electromagnetic Waves

  • Fatemeh Seif,
  • Milad Omidi,
  • Mohamad Reza Bayatiani,
  • Mehdi Ghorbani

DOI
https://doi.org/10.18502/fbt.v9i1.8143
Journal volume & issue
Vol. 9, no. 1

Abstract

Read online

Purpose: The widespread use of mobile phones and Base Transceiver Stations (BTSs) has generated public concern about exposure to Electromagnetic (EM) waves. In this study, the electric field intensity and Specific Absorption Rate (SAR) in the emergency, general hospitalization, radiology, and laboratory departments of four hospitals in Arak (Iran) are reported. Materials and Methods: Electric field strength in the 900 MHz frequency band was obtained using a TES 592 radiometer. Then, SAR induced in the brain, skin, fat and bone tissues were calculated based on equations and the obtained values were compared with the thresholds recommended by the International Commissions. Results: The obtained results showed that the electric field’s mean value was 1.334 V/m which is almost 2.7% of the threshold introduced by the International Commission on Non-Ionizing Radiation Protection (ICNIRP) and 2.6% of the threshold adopted by the Institute of Electrical and Electronics Engineers (IEEE). The highest SAR value was 1.6 W/kg for the skin, which is lower than the threshold values presented by ICNIRP (2 W/kg) and IEEE (1.6 W/kg). Conclusion: The findings of the present work show that for both quantities in Arak hospitals the SAR values are less than the thresholds announced by IEEE and ICNIRP committees. To deal with the concerns of the community that is generally caused by a lack of awareness, the executions of educational and public awareness programs are recommended.

Keywords