BMC Ophthalmology (Sep 2020)

Effect of the electrode array-retina gap distance on visual function in patients with the Argus II retinal prosthesis

  • Abhishek Naidu,
  • Nimra Ghani,
  • Mohammad Saad Yazdanie,
  • Khurram Chaudhary

DOI
https://doi.org/10.1186/s12886-020-01631-6
Journal volume & issue
Vol. 20, no. 1
pp. 1 – 8

Abstract

Read online

Abstract Background Post-implantation visual outcomes in patients with the Argus II Retinal Prosthesis is dependent on a multitude of factors including the positioning of the electrode array on the retina. The purpose of this study is to determine whether the average electrode array-retina gap distance correlates with objective visual function outcomes and sensitivity detection thresholds in patients implanted with the Argus II Retinal Prosthesis. Methods Five patients with implantation of the Argus II Retinal Prosthesis were enrolled in this single-institution retrospective study. Patient demographics were collected from medical records. Visual function data (Square Localization [SL] and Direction of Motion [DOM]) and Optical Coherence Tomography (Cirrus HD-OCT) images were extracted retrospectively from the Argus II Retinal Prosthesis Post-Approval study. Visual function tests were performed with the device OFF and ON at each study visit. Electrode array-retina gap distances were measured at each of the array’s 60 electrodes using the Cirrus HD-OCT software in both the nasotemporal and superoinferior planes. Data was obtained at baseline, and post-operative month 1, month 3, month 6, and year 1. Sensitivity detection thresholds were obtained at the initial programming visit and each reprogramming session. Results Three patients performed significantly better in SL visual function testing with the device ON. Patients that worsened in visual function testing with the device ON in both SL and DOM testing had a statistically significant decrease in performance. The electrode array-retina gap distance was found to effect performance in SL testing in a patient-dependent manner. No effect was found between the electrode-array gap distance and DOM testing or sensitivity detection threshold. Conclusion Our results demonstrate that the electrode array-retina gap distance may affect visual function outcomes in SL testing in certain patients with the Argus II Retinal Prosthesis, and the direction and magnitude of this effect is likely patient-dependent. Furthermore, complete apposition between the electrode array and retina may not always be necessary to achieve optimal visual outcomes.

Keywords