BMC Bioinformatics (Mar 2011)

MRCQuant- an accurate LC-MS relative isotopic quantification algorithm on TOF instruments

  • Haskins William E,
  • Petritis Konstantinos,
  • Zhang Jianqiu

DOI
https://doi.org/10.1186/1471-2105-12-74
Journal volume & issue
Vol. 12, no. 1
p. 74

Abstract

Read online

Abstract Background Relative isotope abundance quantification, which can be used for peptide identification and differential peptide quantification, plays an important role in liquid chromatography-mass spectrometry (LC-MS)-based proteomics. However, several major issues exist in the relative isotopic quantification of peptides on time-of-flight (TOF) instruments: LC peak boundary detection, thermal noise suppression, interference removal and mass drift correction. We propose to use the Maximum Ratio Combining (MRC) method to extract MS signal templates for interference detection/removal and LC peak boundary detection. In our method, MRCQuant, MS templates are extracted directly from experimental values, and the mass drift in each LC-MS run is automatically captured and compensated. We compared the quantification accuracy of MRCQuant to that of another representative LC-MS quantification algorithm (msInspect) using datasets downloaded from a public data repository. Results MRCQuant showed significant improvement in the number of accurately quantified peptides. Conclusions MRCQuant effectively addresses major issues in the relative quantification of LC-MS-based proteomics data, and it provides improved performance in the quantification of low abundance peptides.