Tuberculosis and Respiratory Diseases (Jul 2021)

Measurement of Fractional Exhaled Nitric Oxide in Adults: Comparison of Two Different Analyzers (NIOX VERO and NObreath)

  • Sung-Yoon Kang,
  • Sang Min Lee,
  • Sang Pyo Lee

DOI
https://doi.org/10.4046/trd.2020.0137
Journal volume & issue
Vol. 84, no. 3
pp. 182 – 187

Abstract

Read online

Background Fractional exhaled nitric oxide (FeNO) is a non-invasive marker for eosinophilic airway inflammation and a good predictor of response to corticosteroids. There is a need for a reliable and accurate measurement method, as FeNO measurements have been widely used in clinical practice. Our study aimed to compare two FeNO analyzers and derive a conversion equation for FeNO measurements in adults. Methods We included 99 participants who had chief complaints of chronic cough and difficulty in breathing. The participants underwent concurrent FeNO measurement using NIOX VERO (Circassia AB) and NObreath (Bedfont). We compared the values of the two devices and analyzed their correlation and agreement. We then formulated an equation to convert FeNO values measured by NObreath into those obtained by NIOX VERO. Results The mean age of the participants was 51.2±17.1 years, with a female predominance (58.6%). Approximately 60% of the participants had asthma. The FeNO level measured by NIOX VERO (median, 27; interquartile range [IQR], 15–45) was significantly lower than that measured by NObreath (median, 38; IQR, 22–58; p<0.001). There was a strong positive correlation between the two devices (r=0.779, p<0.001). Additionally, Bland-Altman plots and intraclass correlation coefficient demonstrated a good agreement. Using linear regression, we derived the following conversion equation: natural log (Ln) (NObreath)=0.728×Ln (NIOX VERO)+1.244. Conclusion The FeNO values of NIOX VERO and NObreath were in good agreement and had positive correlations. Our proposed conversion equation could help assess the accuracy of the two analyzers.

Keywords