Ambient dose equivalent measurements in secondary radiation fields at proton therapy facility CCB IFJ PAN in Krakow using recombination chambers
Jakubowska Edyta A.,
Gryziński Michał A.,
Golnik Natalia,
Tulik Piotr,
Stolarczyk Liliana,
Horwacik Tomasz,
Zbroja Katarzyna,
Góra Łukasz
Affiliations
Jakubowska Edyta A.
Institute of Metrology and Biomedical Engineering, Warsaw University of Technology, 8 Św. A. Boboli Str., 02-525 Warsaw, Poland, Tel.: +48 22 234 8643, Fax: +48 22 849 0395
Gryziński Michał A.
National Centre for Nuclear Research, 7 Andrzeja Sołtana Str., 05-400 Otwock/Świerk, Poland
Golnik Natalia
Institute of Metrology and Biomedical Engineering, Warsaw University of Technology, 8 Św. A. Boboli Str., 02-525 Warsaw, Poland, Tel.: +48 22 234 8643, Fax: +48 22 849 0395
Tulik Piotr
Institute of Metrology and Biomedical Engineering, Warsaw University of Technology, 8 Św. A. Boboli Str., 02-525 Warsaw, Poland, Tel.: +48 22 234 8643, Fax: +48 22 849 0395
Stolarczyk Liliana
National Centre for Nuclear Research, 7 Andrzeja Sołtana Str., 05-400 Otwock/Świerk, Poland
Horwacik Tomasz
National Centre for Nuclear Research, 7 Andrzeja Sołtana Str., 05-400 Otwock/Świerk, Poland
Zbroja Katarzyna
National Centre for Nuclear Research, 7 Andrzeja Sołtana Str., 05-400 Otwock/Świerk, Poland
Góra Łukasz
National Centre for Nuclear Research, 7 Andrzeja Sołtana Str., 05-400 Otwock/Świerk, Poland
This work presents recombination methods used for secondary radiation measurements at the Facility for Proton Radiotherapy of Eye Cancer at the Institute for Nuclear Physics, IFJ, in Krakow (Poland). The measurements of H*(10) were performed, with REM-2 tissue equivalent chamber in two halls of cyclotrons AIC-144 and Proteus C-235 and in the corridors close to treatment rooms. The measurements were completed by determination of gamma radiation component, using a hydrogen-free recombination chamber. The results were compared with the measurements using rem meter types FHT 762 (WENDI-II) and NM2 FHT 192 gamma probe and with stationary dosimetric system.