A pilot study for development of a pulmonary function test induction jacket to automate effort in performing the forced vital capacity manoeuvre

  • Prashant Rajdeep,
  • Ketaki Poorey,
  • R. K. Patel,
  • E. R. Oommen

DOI
https://doi.org/10.1038/s41598-023-34930-1
Journal volume & issue
Vol. 13, no. 1
pp. 1 – 11

Abstract

Read online

Abstract The pulmonary function test (PFT) induction jacket was invented to make the process of performing the forced vital capacity (FVC) manoeuvre with a computerized spirometer effortless and productive for both the patient and the medical practitioner. The jacket is composed of three layers of PVC material sealed together to form a single jacket with two chambers. The inner chamber is formed between the inner layer and the middle layer, in which cold water at a temperature of 10 °C is circulated using a connected water unit when triggered. Similarly, the outer chamber is formed between the middle layer and the outer layer, in which air is pressurized using a connected air unit. Thirty volunteers performed the FVC manoeuvre with and without wearing the jacket. There was no difference between the results in spirometry parameters in the participants without a jacket and those with a jacket. However, use of the jacket significantly reduced the number of trials the participants had to undergo to perform spirometry. The jacket automated the FVC manoeuvre by triggering a physiological inspiratory gasp using cold water and circumscribing pressurized air for expiration. Additionally, subsequent advancements in the jacket have been suggested.