BMC Research Notes (Apr 2022)

Usefulness of a commercial LAMP assay for detection of malaria infection, including Plasmodium knowlesi cases, in returning travelers in Spain

  • Alexandra Martín-Ramírez,
  • Marta Lanza,
  • Shamilah Hisam,
  • Ana Perez-Ayala,
  • José M. Rubio

DOI
https://doi.org/10.1186/s13104-022-06037-9
Journal volume & issue
Vol. 15, no. 1
pp. 1 – 6

Abstract

Read online

Abstract Objective Main malaria diagnosis is based on microscopic examination combined with rapid diagnostic tests. Both methods have low sensitivity and specificity. Loop-mediated isothermal amplification techniques have shown a sensitivity similar to PCR but with lower times of performance. This study aimed to assess a commercial LAMP for the diagnosis of malaria (Alethia® Malaria) against the Nested-Multiplex-Malaria PCR, including the analytical sensitivity and the operational characteristics. Results One hundred five samples out of 114 rendered valid results, obtaining 85 positive samples and 18 negative samples with an agreement of 98% compared to the reference method with a sensitivity, specificity and kappa coefficient of 98.84%, 94.74% and 0.94 respectively, with only two discrepant samples. The turnaround time was estimated in 1 h and 30 min, with a cost of 32.67€ per determination. The results showed several advantages of the Alethia ® Malaria, as it was easy to perform, minimal training requirement and 40 min run. Moreover, it includes an internal control to avoid false negatives. However, it also showed some limitations such as the need for a specific amplification and detection device, the detection of only Plasmodium spp. and a very high price.

Keywords