Micromachines (Jan 2023)

Highly Sensitive Electrochemical Endotoxin Sensor Based on Redox Cycling Using an Interdigitated Array Electrode Device

  • Kentaro Ito,
  • Kumi Y. Inoue,
  • Takahiro Ito-Sasaki,
  • Miho Ikegawa,
  • Shinichiro Takano,
  • Kosuke Ino,
  • Hitoshi Shiku

DOI
https://doi.org/10.3390/mi14020327
Journal volume & issue
Vol. 14, no. 2
p. 327

Abstract

Read online

The Limulus amebocyte lysate (LAL) reaction-based assay, the most commonly used endotoxin detection method, requires a skilled technician. In this study, to develop an easy-to-use and highly sensitive endotoxin sensor, we created an electrochemical endotoxin sensor by using an interdigitated array electrode (IDAE) device with advantages of amplifiable signals via redox cycling and portability. We added Boc-Leu-Gly-Arg-p-aminophenol (LGR-pAP) as an electrochemical substrate for an LAL reaction and detected p-aminophenol (pAP) released from LGR-pAP as a product of an endotoxin-induced LAL reaction via an IDAE device. The IDAE device showed a great redox cycling efficiency of 79.8%, and a 4.79-fold signal amplification rate. Then, we confirmed that pAP was detectable in the presence of LGR-pAP through chronoamperometry with the potential of the anode stepped from −0.3 to 0.5 V vs. Ag/AgCl while the cathode was biased at −0.3 V vs. Ag/AgCl. Then, we performed an endotoxin assay by using the IDAE device. Our endotoxin sensor detected as low as 0.7 and 1.0 endotoxin unit/L after the LAL reaction for 1 h and 45 min, respectively, and these data were within the cut-off value for ultrapure dialysis fluid. Therefore, our highly sensitive endotoxin sensor is useful for ensuring medical safety.

Keywords