Sensors (Jan 2023)

Reliability and Validity of a Wearable Sensing System and Online Gait Analysis Report in Persons after Stroke

  • Anne Schwarz,
  • Adib Al-Haj Husain,
  • Lorenzo Einaudi,
  • Eva Thürlimann,
  • Julia Läderach,
  • Chris Awai Easthope,
  • Jeremia P. O. Held,
  • Andreas R. Luft

DOI
https://doi.org/10.3390/s23020624
Journal volume & issue
Vol. 23, no. 2
p. 624

Abstract

Read online

The restoration of gait and mobility after stroke is an important and challenging therapy goal due to the complexity of the potentially impaired functions. As a result, precise and clinically feasible assessment methods are required for personalized gait rehabilitation after stroke. The aim of this study is to investigate the reliability and validity of a sensor-based gait analysis system in stroke survivors with different severities of gait deficits. For this purpose, 28 chronic stroke survivors (9 women, ages: 62.04 ± 11.68 years) with mild to moderate walking impairments performed a set of ambulatory assessments (3× 10MWT, 1× 6MWT per session) twice while being equipped with a sensor suit. The derived gait reports provided information about speed, step length, step width, swing and stance phases, as well as joint angles of the hip, knee, and ankle, which we analyzed for test-retest reliability and hypothesis testing. Further, test-retest reliability resulted in a mean ICC of 0.78 (range: 0.46–0.88) for walking 10 m and a mean ICC of 0.90 (range: 0.63–0.99) for walking 6 min. Additionally, all gait parameters showed moderate-to-strong correlations with clinical scales reflecting lower limb function. These results support the applicability of this sensor-based gait analysis system for individuals with stroke-related walking impairments.

Keywords