EJNMMI Physics (Jun 2020)

Influence of sampling schedules on [177Lu]Lu-PSMA dosimetry

  • Andreas Rinscheid,
  • Peter Kletting,
  • Matthias Eiber,
  • Ambros J. Beer,
  • Gerhard Glatting

DOI
https://doi.org/10.1186/s40658-020-00311-0
Journal volume & issue
Vol. 7, no. 1
pp. 1 – 14

Abstract

Read online

Abstract Background Individualized dosimetry is recommended for [177Lu]Lu-PSMA radioligand therapy (RLT) which is resource-intensive and protocols are often not optimized. Therefore, a simulation study was performed focusing on the determination of efficient optimal sampling schedules (OSS) for renal and tumour dosimetry by investigating different numbers of time points (TPs). Methods Sampling schedules with 1–4 TPs were investigated. Time-activity curves of the kidneys and two tumour lesions were generated based on a physiologically based pharmacokinetic (PBPK) model and biokinetic data of 13 patients who have undergone [177Lu]Lu-PSMA I&T therapy. Systematic and stochastic noise of different ratios was considered when modelling time-activity data sets. Time-integrated activity coefficients (TIACs) were estimated by simulating the hybrid planar/SPECT method for schedules comprising at least two TPs. TIACs based on one single SPECT/CT measurement were estimated using an approximation for reducing the number of fitted parameters. For each sampling schedule, the root-mean-squared error (RMSE) of the deviations of the simulated TIACs from the ground truths for 1000 replications was used as a measure for accuracy and precision. Results All determined OSS included a late measurement at 192 h p.i., which was necessary for accurate and precise tumour TIACs. OSS with three TPs were identified to be 3–4, 96–100 and 192 h with an additional SPECT/CT measurement at the penultimate TP. Kidney and tumour RMSE of 6.4 to 7.7% and 6.3 to 7.8% were obtained, respectively. Shortening the total time for dosimetry to e.g. 96 h resulted in kidney and tumour RMSE of 6.8 to 8.3% and 9.1 to 11%, respectively. OSS with four TPs showed similar results as with three TPs. Planar images at 4 and 68 h and a SPECT/CT shortly after the 68 h measurement led to kidney and tumour RMSE of 8.4 to 12% and 12 to 16%, respectively. One single SPECT/CT measurement at 52 h yielded good approximations for the kidney TIACs (RMSE of 7.0%), but led to biased tumour TIACs. Conclusion OSS allow improvements in accuracy and precision of renal and tumour dosimetry for [177Lu]Lu-PSMA therapy with potentially less effort. A late TP is important regarding accurate tumour TIACs.

Keywords