A Novel Equivalent Time Sampling-Based Method for Pulse Transit Time Estimation with Applications into the Cardiovascular Disease Diagnosis

  • Giorgia Fiori,
  • Fabio Fuiano,
  • Silvia Conforto,
  • Salvatore Andrea Sciuto,
  • Andrea Scorza

DOI
https://doi.org/10.3390/s23115005
Journal volume & issue
Vol. 23, no. 11
p. 5005

Abstract

Read online

The increasing incidence of cardiovascular diseases (CVDs) is reflected in additional costs for healthcare systems all over the world. To date, pulse transit time (PTT) is considered a key index of cardiovascular health status and for diagnosis of CVDs. In this context, the present study focuses on a novel image analysis-based method for PTT estimation through the application of equivalent time sampling. The method, which post-processes color Doppler videos, was tested on two different setups: a Doppler flow phantom set in pulsatile mode and an in-house arterial simulator. In the former, the Doppler shift was due to the echogenic properties of the blood mimicking fluid only, since the phantom vessels are non-compliant. In the latter, the Doppler signal relied on the wall movement of compliant vessels in which a fluid with low echogenic properties was pumped. Therefore, the two setups allowed the measurement of the flow average velocity (FAV) and the pulse wave velocity (PWV), respectively. Data were collected through an ultrasound diagnostic system equipped with a phased array probe. Experimental outcomes confirm that the proposed method can represent an alternative tool for the local measurement of both FAV in non-compliant vessels and PWV in compliant vessels filled with low echogenic fluids.