Journal of Cardiothoracic Surgery (Nov 2018)

Epicardial infrared ablation to create a linear conduction block on a beating right atrium

  • Hiroshi Kubota,
  • Hidehito Endo,
  • Hikaru Ishii,
  • Hiroshi Tsuchiya,
  • Yusuke Inaba,
  • Yu Takahashi,
  • Katsunari Terakawa

DOI
https://doi.org/10.1186/s13019-018-0801-y
Journal volume & issue
Vol. 13, no. 1
pp. 1 – 6

Abstract

Read online

Abstract Background It is still difficult to create a secure linear conduction block on a beating heart from the epicardial side. To overcome this drawback we developed an infrared coagulator equipped with a cuboid light-guiding quartz rod. This study was designed to electrophysiologically confirm the efficacy of a new ablation probe using infrared energy in a clinical case. Methods The infrared light from a lamp is focused into the newly developed cuboid quartz rod, which has a rectangular distal exit-plane that allows 30 mm × 10 mm linear photocoagulation. Two pairs of electrodes were attached to the right atrium of a patient who was undergoing surgery. Each pair of electrodes was placed 10 mm from an ablation line. The change in conduction time between the two pairs of electrodes was measured during ablation. The predicted conduction time delay ratio was 1.54. Results The actual conduction time after ablation was 1.38–1.43 times longer than the pre-ablation conduction time. Conclusions The infrared ablation using a newly developed cuboid probe made it possible to create a linear conduction block on the beating right atrial free wall clinically.