Iranian Journal of Parasitology (Sep 2020)

Evaluation of the Diagnostic Performance of Recombinant Antigen B1 for Detection of Cystic Echinococcosis Using Lateral Flow Dipstick Test

  • Rahmah NOORDIN,
  • Sam KHANBABAIE,
  • Muhammad HAFIZNUR YUNUS,
  • Hanspeter MARTI,
  • Beatrice NICKEL,
  • Majid FASIHI HARANDI,
  • Saeid NASIBI

DOI
https://doi.org/10.18502/ijpa.v15i3.4191
Journal volume & issue
Vol. 15, no. 3
pp. 290 – 298

Abstract

Read online

Background: Human echinococcosis is a neglected zoonotic disease distributed worldwide. It comprises cystic and alveolar forms, the former being the more prevalent disease. Imaging techniques are the first choice for diagnosis of cystic echinococcosis and serology is used as an additional diagnostic technique in doubtful cases or as the sole test in low-resource settings. Rapid diagnostic tests are useful and convenient for immunodiagnosis of cystic echinococcosis in endemic areas, where medical facilities often struggle with limited resources. Methods: Recently, we have developed Hyd Rapid™, an IgG4 lateral flow dipstick test using recombinant antigen B1 for detection of cystic echinococcosis. This study was performed between 2016 until 2018 at the Institute for Research in Molecular Medicine, Universiti Sains Malaysia. The diagnostic performance of Hyd Rapid™ was tested in-house and at two international laboratories in Switzerland and Iran. Results: The overall diagnostic sensitivity for detection of cystic and alveolar echinococcosis was 95% (56/59). Meanwhile, the diagnostic specificity, with and without exclusion of cysticercosis and fascioliasis, was 100% (n=48) and 88% (63/72), respectively. Conclusion: Hyd Rapid™ detected cystic echinococcosis as well as probable cases of alveolar echinococcosis. Therefore, Hyd Rapid™ showed good potential as a serological tool for echinococcosis, and merits further evaluation.

Keywords