Sensors (Jun 2022)

Electric-Field-Based Guidance for Percutaneous Catheter Vessel Crossing

  • Mamadou Diallo,
  • Clemens Eder,
  • Daniel Brasier,
  • Sorin Popa,
  • Robert Dickinson

DOI
https://doi.org/10.3390/s22134928
Journal volume & issue
Vol. 22, no. 13
p. 4928

Abstract

Read online

Percutaneous procedures to divert blood flow from one blood vessel to another can be performed with intravascular catheters but demand a method to align a crossing needle from one vessel to another. Fluoroscopic imaging alone is not adequate, and it is preferable to have a sensor on one catheter that detects the correct alignment of an incoming needle. This can be implemented by generating dipole electric fields from the crossing catheter which are detected by a receiving catheter in the target vessel and, thus, can calculate and display the degree of alignment, permitting the operator to rotate the crossing catheter to guarantee alignment when deploying a crossing needle. Catheters were built using this concept and evaluated in vitro. The results show that accurate alignment is achieved, and a successful crossing can be made. The concept is being further developed for further clinical evaluation.

Keywords