Research on Construction of Test Environment for Assessment of RF-Induced Heating Effects of Implants

  • Yonghua LI,
  • Lingfeng LU,
  • Jing WANG,
  • Chengling LI,
  • Pengfei YANG,
  • Zhichao WANG,
  • Dajing WU,
  • Xun LIU,
  • Sheng HU

DOI
https://doi.org/10.12455/j.issn.1671-7104.240031
Journal volume & issue
Vol. 48, no. 3
pp. 281 – 284

Abstract

Read online

In magnetic resonance examination, the interaction between implants and the radio frequency (RF) fields induces heating in human tissue and may cause tissue damage. To assess the RF-induced heating of implants, three steps should be executed, including electromagnetic model construction, electromagnetic model validation, and virtual human body simulations. The crucial step of assessing RF-induced heating involves the construction of a test environment for electromagnetic model validation. In this study, a hardware environment, comprised of a RF generation system, electromagnetic field measurement system, and a robotic arm positioning system, was established. Furthermore, an automated control software environment was developed using a Python-based software development platform to enable the creation of a high-precision automated integrated test environment. The results indicate that the electric field generated in this test environment aligns well with the simulated electric field, making it suitable for assessing the RF-induced heating effects of implants.