CVIR Endovascular (Jan 2020)

Recanalisation of an axillary vein occlusion jailed by a migrated cephalic arch stent-graft using the TruePath chronic total occlusion drilling device

  • Ankur Patel,
  • Shaun Xavier Ju Min Chan,
  • Kun Da Zhuang

DOI
https://doi.org/10.1186/s42155-020-0098-5
Journal volume & issue
Vol. 3, no. 1
pp. 1 – 5

Abstract

Read online

Abstract Background Stent placement in the cephalic arch is being used with increasing frequency. Late complications of bare metal and stent grafts in dialysis access, in particular stent migration, are often under-reported and can lead to compromise of future dialysis circuits. Case presentation A 52-year-old man developed acute arm swelling 2 days after creation of a left arm brachio-basilic arteriovenous graft. The axillary vein was found to be jailed by a previously deployed cephalic arch stent graft which had migrated into the subclavian vein. There was failure to cross through the fabric of the stent graft using conventional chronic total occlusion wires and techniques. A TruePath device was used successfully to cross through the fabric of migrated cephalic arch stent graft and recanalise the short subclavian-axillary vein occlusion. Conclusion The adapted use of a drilling chronic total occlusion device to drill through the fabric of migrated stent graft was performed successfully to allow complete recanalisation of the occluded axillary vein.