Scientific Reports (Sep 2024)

Evaluation of repeatability and agreement of two optical biometers for intraocular lens power calculation

  • Irene Martinez Alberquilla,
  • Saga Svensson,
  • Javier Ruiz-Alcocer,
  • David Madrid-Costa,
  • Alberto Dominguez-Vicent,
  • Abinaya Priya Venkataraman

DOI
https://doi.org/10.1038/s41598-024-73206-0
Journal volume & issue
Vol. 14, no. 1
pp. 1 – 8

Abstract

Read online

Abstract The repeatability of two biometers (Lenstar-LS900 and Eyestar-900) to measure ocular parameters and intraocular lens (IOL) power calculation, and their agreement were evaluated. 134 eyes of 134 participants were measured thrice with each biometer. Axial length (AL), anterior chamber depth (ACD), lens thickness (LT) and keratometry (K) were evaluated. The IOL power was calculated using different formulas. The repeatability limit (RLimit), the mean differences (MD) and the limits of agreement (LoA) were calculated. The RLimits for all parameters were higher with Lenstar compared to Eyestar. RLimits were lower than 0.50 D except for Barrett Universal II (0.54 D) and Haigis (0.51 D) formulas with the Lenstar. Mean differences were lower than 0.01 mm for AL, ACD and LT, and lower than 0.03 D for K. MD ranged from 0 to 0.02 D for all formulas except for Barrett and Hill. When dividing the sample into subgroups (short, normal and long eyes), the MDs were similar for the IOL power and were lower than 0.03 D, except for the Barrett and Hill formulas. Both biometers provide repeatable biometry and IOL power calculations. The LoA interval for the IOL power calculation was between 0.75 and 1.50D, which was similar among the subgroups.