BMC Medical Imaging (Jun 2024)

Comparison of 96-kV and 120-kV cone-beam CT for the assessment of cochlear implants

  • Iris Burck,
  • Ibrahim Yel,
  • Simon Martin,
  • Moritz H. Albrecht,
  • Vitali Koch,
  • Christian Booz,
  • Daniel Pinto dos Santos,
  • Benjamin Kaltenbach,
  • Hanns Ackermann,
  • Juha Koivisto,
  • Silke Helbig,
  • Timo Stöver,
  • Thomas J. Vogl,
  • Jan-Erik Scholtz

DOI
https://doi.org/10.1186/s12880-024-01322-4
Journal volume & issue
Vol. 24, no. 1
pp. 1 – 10

Abstract

Read online

Abstract Background To compare the diagnostic value of 120-kV with conventional 96-kV Cone-Beam CT (CBCT) of the temporal bone after cochlear implant (CI) surgery. Methods This retrospective study included CBCT scans after CI surgery between 06/17 and 01/18. CBCT allowed examinations with 96-kV or 120-kV; other parameters were the same. Two radiologists independently evaluated following criteria on 5-point Likert scales: osseous spiral lamina, inner and outer cochlear wall, semi-circular canals, mastoid trabecular structure, overall image quality, metal and motion artefacts, depiction of intracochlear electrode position and visualisation of single electrode contacts. Effective radiation dose was assessed. Results Seventy-five patients (females, n = 39 [52.0%], mean age, 55.8 ± 16.5 years) were scanned with 96-kV (n = 32, 42.7%) and 120-kV (n = 43, 57.3%) protocols including CI models from three vendors (vendor A n = 7; vendor B n = 43; vendor C n = 25). Overall image quality, depiction of anatomical structures, and electrode position were rated significantly better in 120-kV images compared to 96-kV (all p < = 0.018). Anatomical structures and electrode position were rated significantly better in 120-kV CBCT for CI models from vendor A and C, while 120-kV did not provide improved image quality in CI models from vendor B. Radiation doses were significantly higher for 120-kV scans compared to 96-kV (0.15 vs. 0.08 mSv, p < 0.001). Conclusions 120-kV and 96-kV CBCT provide good diagnostic images for the postoperative CI evaluation. While 120-kV showed improved depiction of temporal bone and CI electrode position compared to 96-kV in most CI models, the 120-kV protocol should be chosen wisely due to a substantially higher radiation exposure.

Keywords