Current Directions in Biomedical Engineering (Sep 2015)

Heart rate monitoring in ultra-high-field MRI using frequency information obtained from video signals of the human skin compared to electrocardiography and pulse oximetry

  • Spicher Nicolai,
  • Maderwald Stefan,
  • Ladd Mark E.,
  • Kukuk Markus

DOI
https://doi.org/10.1515/cdbme-2015-0018
Journal volume & issue
Vol. 1, no. 1
pp. 69 – 72

Abstract

Read online

Videos of the human skin contain subtle color variations associated with the blood volume pulse. This remote photoplethysmography signal can be used for heart rate monitoring and represents an alternative to signals obtained from contact-based hardware. We developed an algorithm that estimates the heart rate in real-time from photoplethysmography signals and evaluate its performance in the context of ultra-high-field magnetic resonance imaging. We compare its accuracy to heart rate values estimated from electrocardiography and finger pulse oximetry triggers, obtained from MR vendor-provided hardware. For eight subjects, two experiments are conducted with the patient table outside and inside a 7 Tesla scanner. During both 5 min setups, heart rates from the algorithm and contact-based methods are stored. Their comparison suggests technical feasibility of the contactless method but that it is inferior in accuracy compared to contact-based hardware and that low heart rates (≤50 beats per minute) and adequate illumination are major challenges for practical feasibility.