EPJ Web of Conferences (Jan 2018)

Characterization and localization of partial-discharge-induced pulses in fission chambers designed for sodium-cooled fast reactors

  • Galli G.,
  • Hamrita H.,
  • Jammes C.,
  • Kirkpatrick M.J.,
  • Odic E.,
  • Dessante Ph.,
  • Molinie Ph.,
  • Cantonnet B.,
  • Nappé J-C.

DOI
https://doi.org/10.1051/epjconf/201817003002
Journal volume & issue
Vol. 170
p. 03002

Abstract

Read online

During the operation of the Superphenix and Phenix reactors, an aberrant electrical signal was detected from the fission chambers used for neutron flux monitoring. This signal, thought to be due to partial electrical discharge (PD) is similar to the signal resulting from neutron interactions, and is generated in fission chambers at temperatures above 400 °C. This paper reports work on the characterization and localization of the source of this electrical signal in a High Temperature Fission Chamber (HTFC). The relation between the shape of the PD signal and various parameters (nature and pressure of the chamber filling gas, electrode gap distance, and fission chamber geometry) are first described. Next, experiments designed to identify the location within the chambers where the PD are being generated are presented. After verification and refinement of the results of these localization studies, it should be possible to propose changes to the fission chamber in order to reduce or eliminate the PD signal.

Keywords