Technology in Cancer Research & Treatment (Oct 2020)

Performance of Automated Dissection on Formalin-Fixed Paraffin-Embedded Tissue Sections for the 21-Gene Recurrence Score Assay

  • Peng Qi MD,
  • Qian-ming Bai MD,
  • Qian-lan Yao MD,
  • Wen-tao Yang MD,
  • Xiao-yan Zhou MD

DOI
https://doi.org/10.1177/1533033820960760
Journal volume & issue
Vol. 19

Abstract

Read online

This study aimed to compare the performance of MilliSect dissection and manual dissection. Twenty-five formalin-fixed paraffin-embedded (FFPE) breast cancer tissue blocks were selected for comparison. Specific areas of interest (AOIs) in invasive carcinoma on tissue sections were transferred to dissection slides by manual macrodissection or the MilliSect instrument. The comparison criteria were 1) the time required for dissection; 2) RNA concentration and purity; 3) RNA quantity of 5 housekeeping genes (by RT-qPCR); and 4) ER, PR, HER2, Ki-67 and recurrence score (RS) values (by the 21-gene assay). Then, tumor-adjacent tissues, including fibrocollagenous and epithelial tissues, from the same selected tissue blocks of 8 of 25 patients were scraped using the mesodissection method, and their RS values were assessed to evaluate the influence of tumor-adjacent tissues on the target AOIs. Ultimately, 4 AOIs of invasive ductal carcinoma (IDC) from 1 tissue block of another 4 patients with lymph node (LN) metastases each, LN tissue and a mixture of IDC and LN tissue from the other tissue block of the same 4 patients were mesodissected to evaluate the influence of infiltrating lymphocyte levels on the RS values of AOIs. In our experience, the MilliSect instrument, which provides process management documentation, required more time than manual macrodissection (on average, approximately 9.1 min per sample versus 5.8 min per sample, respectively). The RNA yield and quality of the dissected tissues were comparable for the 2 methods. However, the tumor-adjacent tissues of the AOIs may influence the RS to some extent. Tumor-infiltrating lymphocytes (TILs) can dramatically increase RSs, far exceeding the influence of tumor-adjacent fibrocollagenous and epithelial tissues. In conclusion, MilliSect mesodissection is comparable to manual dissection. This mesodissection tool may facilitate AOI alignment and the dissection process for the 21-gene RS assay. Samples whose adjacent tissues are intermixed with TILs warrant special attention.