Applied Sciences (Nov 2023)
Application of Performance Test Method in Korea for LED Optical Medical Device Samples
Abstract
This study obtained key performance items related to phototherapy and the standards of light-emitting diode (LED) phototherapy devices through the published literature to secure LED phototherapy device safety in South Korea. Based on these items and IEC 62471-1 for the photobiological safety of lamps and lamp systems, a performance test method and wavelength parameters, which are the main performance indicators of the LED phototherapy device, were derived. We conducted a performance test derived from this study using samples with the same performance as the sample that is being developed. The samples comprised a laser and IR LED, red LED, and blue LED. All samples were within ±5%, of the standard threshold. The 455 nm blue LED and 625 nm red LED were close to the set value with a difference of 1 to 2 nm; their wavelengths’ accuracy was within ±0.32%, ±0%, and ±0.219%, depending on the samples’ wavelengths, which was the same or similar to the error range of ±0%. As for laser and IR LED, the difference from the set value was as large as 2 to 7 nm, and their error ranges were close to ±1%, as shown by ±0.615%, ±0.828%, or ±0.591%, which were larger than the values for red and blue LEDs.