eJournal of Oral Maxillofacial Research (Oct 2013)

Developing a New Appliance to Dissipate Mechanical Load on Teeth and Improve Limitation of Vertical Mouth Opening

  • Takashi Satomi,
  • Toshiaki Tanaka,
  • Takehito Kobayashi,
  • Mituyoshi Iino

DOI
https://doi.org/10.5037/jomr.2013.4304
Journal volume & issue
Vol. 4, no. 3
p. e4

Abstract

Read online

Objectives: The principle of leverage to superpose the convex surfaces of two shells was applied to develop a device for treating limitation of mouth opening and called it the “shell-shaped mouth opener” and analyzed pressure on the teeth with the TheraBite® appliance and the shell-shaped mouth opening appliance. Material and Methods: To compare the TheraBite® appliance and the shell-shaped mouth opening appliance, pressure on the teeth in the dentition model with both devices was analyzed using the Inastomer® flexible conductive sensor. Results: The load was better dispersed to each tooth in the shell-shaped mouth opening appliance in the all quadrants compared to the TheraBite® appliance. Conclusions: The present study revealed that the shell-shaped mouth opening appliance which was originally invented in our lab, dissipated the mechanical load on teeth more evenly than the TheraBite® appliance.