Annales Geophysicae (Mar 2006)

Electric field measurements on Cluster: comparing the double-probe and electron drift techniques

  • A. I. Eriksson,
  • M. André,
  • B. Klecker,
  • H. Laakso,
  • P.-A. Lindqvist,
  • F. Mozer,
  • G. Paschmann,
  • G. Paschmann,
  • A. Pedersen,
  • J. Quinn,
  • R. Torbert,
  • K. Torkar,
  • H. Vaith,
  • H. Vaith

DOI
https://doi.org/10.5194/angeo-24-275-2006
Journal volume & issue
Vol. 24
pp. 275 – 289

Abstract

Read online

The four Cluster satellites each carry two instruments designed for measuring the electric field: a double-probe instrument (EFW) and an electron drift instrument (EDI). We compare data from the two instruments in a representative sample of plasma regions. The complementary merits and weaknesses of the two techniques are illustrated. EDI operations are confined to regions of magnetic fields above 30 nT and where wave activity and keV electron fluxes are not too high, while EFW can provide data everywhere, and can go far higher in sampling frequency than EDI. On the other hand, the EDI technique is immune to variations in the low energy plasma, while EFW sometimes detects significant nongeophysical electric fields, particularly in regions with drifting plasma, with ion energy (in eV) below the spacecraft potential (in volts). We show that the polar cap is a particularly intricate region for the double-probe technique, where large nongeophysical fields regularly contaminate EFW measurments of the DC electric field. We present a model explaining this in terms of enhanced cold plasma wake effects appearing when the ion flow energy is higher than the thermal energy but below the spacecraft potential multiplied by the ion charge. We suggest that these conditions, which are typical of the polar wind and occur sporadically in other regions containing a significant low energy ion population, cause a large cold plasma wake behind the spacecraft, resulting in spurious electric fields in EFW data. This interpretation is supported by an analysis of the direction of the spurious electric field, and by showing that use of active potential control alleviates the situation.