Medicina (Sep 2022)

Accuracy of the Tibial Component Alignment by Extramedullary System Using Simple Radiographic References in Total Knee Arthroplasty

  • Jin-Ho Cho,
  • Jun Young Choi,
  • Sung-Sahn Lee

DOI
https://doi.org/10.3390/medicina58091212
Journal volume & issue
Vol. 58, no. 9
p. 1212

Abstract

Read online

Background and Objectives: The tibial component alignment is an important issue for the longevity of total knee arthroplasty (TKA). The purpose of our study was to investigate the usefulness of proximal tibial references determined by pre-operative radiography and intraoperative C-arm-guided hip and ankle center marking for the extramedullary guided tibial cut in mild (Materials and Methods: A total of 150 consecutive patients (220 cases) who underwent total knee arthroplasty who were recruited from July 2011 to April 2017 were reviewed retrospectively. Before surgery, the proximal tibial reference point and medio-lateral cut thickness difference were identified. Then, hip and ankle centers were checked using a C-arm intensifier intraoperatively. The hip–knee–ankle (HKA) alignment and medial proximal tibial angle (MPTA) were assessed pre-operatively and post-operatively. More than 3° varus or valgus of HKA alignment or tibial component angle was defined as an outlier. Results: Mean follow-up duration was 26.9 months. Among 220 cases, 111 cases are classified as mild varus group and 109 cases are classified as severe varus group. The HKA alignment is significantly improved (p p = 0.62). Conclusion: Measurement of proximal tibial radiographic references and checking the C-arm-guided intraoperative hip and ankle center could be helpful to obtain the favorable coronal position of the tibial component in the extramedullary guided tibial cut.

Keywords