Scientific Reports (Nov 2022)
A formula to improve the reliability of optical axial length measurement in IOL power calculation
Abstract
Abstract To verify the influence of axial length (AL) variations after cataract surgery in IOL power calculation. Patients underwent ophthalmic evaluation before surgery, including optical biometry with IOLMaster 500. Same exams were repeated 2 months after surgery: AL of operated eye was evaluated using two modes (pseudophakic/aphakic options). Mean Keratometry and AL changes were analyzed. Furthermore, refractive prediction error (PE) was back-calculated with Barrett Universal-II, Hoffer-Q, Holladay-1 and SRK/T formulas. To eliminate any systematic error, the mean error (ME) was zeroed-out for each formula. MEs and median absolute errors (MedAEs) of PEs were analyzed. Two-hundred-one operated eyes of 201 patients and 201 opposite eyes were evaluated. In operated eyes, mean AL difference was − 0.11 ± 0.07 mm (p 0.05). Instead, only MEs of PEs obtained with postoperative ALs-pseudophakic option were not-statistically different from zero (p > 0.05). AL measurement change after cataract surgery is probably due to a systematic error in optical biometer in case of phakic eyes. A correction factor applied to preoperative AL could eliminate any systematic error in IOL power calculation without modifying the lens constant.