PLoS ONE (Jan 2018)

Efficacy of a one-catheter concept for transradial coronary angiography.

  • Christoph Langer,
  • Julia Riehle,
  • Helge Wuttig,
  • Stephanie Dürrwald,
  • Helmut Lange,
  • Alexander Samol,
  • Norbert Frey,
  • Marcus Wiemer

DOI
https://doi.org/10.1371/journal.pone.0189899
Journal volume & issue
Vol. 13, no. 1
p. e0189899

Abstract

Read online

Transradial coronary angiography (TRC) can be performed with a one-catheter approach for the right and left coronary ostium (R/LCO). We investigated the performance of a special diagnostic catheter widely used for the one-catheter-approach, the Tiger (Tiger II, TerumoTM).In a dual center registry we analyzed 1412 TRC-procedures exclusively performed by experienced TRC-operators. We compared the performance of the Tiger with Judkins catheters by retrospectively judging ostial catheter stability during contrast injection, and by measuring contrast use, fluoroscopy time (FT) and complication rate.Poor or failed ostial engagement was found in 40.5% in the Tiger group, compared to 46.6% with the use of Judkins catheters (p<0.183). Ostial instability of the Tiger was found more often during engagement of the LCO than the RCO (34.4% vs. 10.8%, p<0.001), whereas it was similar in the LCO and RCO for Judkins catheters (27.4% vs. 26.7%, p = 0.840). TRC-procedures performed with Tiger catheters were associated with less contrast volume (63.48 ± 29.83mL vs. 82.51 ± 56.58mL, p<0.004) and shorter FT than with Judkins catheters. (198.27 ± 194.8sec vs. 326.85 ± 329.70sec). Forearm hematomas occurred less often with the Tiger (1.2% vs. 6.6%, p< 0.02).The Tiger employed as a single catheter in TRC is an effective tool to achieve lower contrast volume and fluoroscopy time at a low complication rate. Unstable engagement affects predominantly the left coronary artery, but its overall frequency is similar for both, the Tiger and Judkins catheters.