PLoS ONE (Jan 2023)

Head-to-head comparison of nasal and nasopharyngeal sampling using SARS-CoV-2 rapid antigen testing in Lesotho.

  • Niklaus D Labhardt,
  • Lucia González Fernández,
  • Bulemba Katende,
  • Josephine Muhairwe,
  • Moniek Bresser,
  • Alain Amstutz,
  • Tracy R Glass,
  • Morten Ruhwald,
  • Jilian A Sacks,
  • Camille Escadafal,
  • Mathabo Mareka,
  • Sekhele M Mooko,
  • Margaretha de Vos,
  • Klaus Reither

DOI
https://doi.org/10.1371/journal.pone.0278653
Journal volume & issue
Vol. 18, no. 3
p. e0278653

Abstract

Read online

ObjectivesTo assess the real-world diagnostic performance of nasal and nasopharyngeal swabs for SD Biosensor STANDARD Q COVID-19 Antigen Rapid Diagnostic Test (Ag-RDT).MethodsIndividuals ≥5 years with COVID-19 compatible symptoms or history of exposure to SARS-CoV-2 presenting at hospitals in Lesotho received two nasopharyngeal and one nasal swab. Ag-RDT from nasal and nasopharyngeal swabs were performed as point-of-care on site, the second nasopharyngeal swab used for polymerase chain reaction (PCR) as the reference standard.ResultsOut of 2198 participants enrolled, 2131 had a valid PCR result (61% female, median age 41 years, 8% children), 84.5% were symptomatic. Overall PCR positivity rate was 5.8%. The sensitivity for nasopharyngeal, nasal, and combined nasal and nasopharyngeal Ag-RDT result was 70.2% (95%CI: 61.3-78.0), 67.3% (57.3-76.3) and 74.4% (65.5-82.0), respectively. The respective specificity was 97.9% (97.1-98.4), 97.9% (97.2-98.5) and 97.5% (96.7-98.2). For both sampling modalities, sensitivity was higher in participants with symptom duration ≤ 3days versus ≤ 7days. Agreement between nasal and nasopharyngeal Ag-RDT was 99.4%.ConclusionsThe STANDARD Q Ag-RDT showed high specificity. Sensitivity was, however, below the WHO recommended minimum requirement of ≥ 80%. The high agreement between nasal and nasopharyngeal sampling suggests that for Ag-RDT nasal sampling is a good alternative to nasopharyngeal sampling.