PLoS ONE (Jan 2021)

Three step flow focusing enables image-based discrimination and sorting of late stage 1 Haematococcus pluvialis cells.

  • Daniel Kraus,
  • Andreas Kleiber,
  • Enrico Ehrhardt,
  • Matthias Leifheit,
  • Peter Horbert,
  • Matthias Urban,
  • Nils Gleichmann,
  • Günter Mayer,
  • Jürgen Popp,
  • Thomas Henkel

DOI
https://doi.org/10.1371/journal.pone.0249192
Journal volume & issue
Vol. 16, no. 3
p. e0249192

Abstract

Read online

Label-free and gentle separation of cell stages with desired target properties from mixed stage populations are a major research task in modern biotechnological cultivation process and optimization of micro algae. The reported microfluidic sorter system (MSS) allows the subsequent investigation of separated subpopulations. The implementation of a viability preserving MSS is shown for separation of late stage 1 Haematococcus pluvialis (HP) cells form a mixed stage population. The MSS combines a three-step flow focusing unit for aligning the cells in single file transportation mode at the center of the microfluidic channel with a pure hydrodynamic sorter structure for cell sorting. Lateral displacement of the cells into one of the two outlet channels is generated by piezo-actuated pump chambers. In-line decision making for sorting is based on a user-definable set of image features and properties. The reported MSS significantly increased the purity of target cells in the sorted population (94%) in comparison to the initial mixed stage population (19%).