Medicina (Feb 2024)

A Novel Method for the Measurement of the Vaginal Wall Thickness by Transvaginal Ultrasound: A Study of Inter- and Intra-Observer Reliability

  • Sara Bosio,
  • Marta Barba,
  • Annalisa Vigna,
  • Alice Cola,
  • Desirèe De Vicari,
  • Clarissa Costa,
  • Silvia Volontè,
  • Matteo Frigerio

DOI
https://doi.org/10.3390/medicina60030370
Journal volume & issue
Vol. 60, no. 3
p. 370

Abstract

Read online

Background and Objectives: A consensus regarding the optimal sonographic technique for measuring vaginal wall thickness (VWT) is still absent in the literature. This study aims to validate a new method for measuring VWT using a biplanar transvaginal ultrasound probe and assess both its intra-operator and inter-operator reproducibility. Material and Methods: This prospective study included patients with genitourinary syndrome of menopause-related symptoms. Women were scanned using a BK Medical Flex Focus 400 with the 65 × 5.5 mm linear longitudinal transducer of an endovaginal biplanar probe (BK Medical probe 8848, BK Ultrasound, Peabody, MA, USA). Vaginal wall thickness (VWT) measurements were acquired from the anterior and posterior vaginal wall at three levels. Results: An inter-observer analysis revealed good consistency between operators at every anatomical site, and the intra-class coefficient ranged from 0.931 to 0.987, indicating high reliability. An intra-observer analysis demonstrated robust consistency in vaginal wall thickness measurements, with an intra-class coefficient exceeding 0.9 for all anatomical sites. Conclusions: The measurement of vaginal wall thickness performed by transvaginal biplanar ultrasound was easy and demonstrated good intra- and inter-operator reliability.

Keywords