AIP Advances (Jan 2018)

A simplified focusing and astigmatism correction method for a scanning electron microscope

  • Yihua Lu,
  • Xianmin Zhang,
  • Hai Li

DOI
https://doi.org/10.1063/1.5009683
Journal volume & issue
Vol. 8, no. 1
pp. 015124 – 015124-10

Abstract

Read online

Defocus and astigmatism can lead to blurred images and poor resolution. This paper presents a simplified method for focusing and astigmatism correction of a scanning electron microscope (SEM). The method consists of two steps. In the first step, the fast Fourier transform (FFT) of the SEM image is performed and the FFT is subsequently processed with a threshold to achieve a suitable result. In the second step, the threshold FFT is used for ellipse fitting to determine the presence of defocus and astigmatism. The proposed method clearly provides the relationships between the defocus, the astigmatism and the direction of stretching of the FFT, and it can determine the astigmatism in a single image. Experimental studies are conducted to demonstrate the validity of the proposed method.